UNO-T-H-Premixed or CO2: all-in-one stage top incubator's controller with for application with pre-mixed 5% CO2 gas or 100% CO2 and background Air, that includes:
•Up to 6 independent temperature channels including
* Uno Stage Top Temperature Controller
* Humidity module temperature control: 90% ca through temperature controlled gas bubbling column
* Temperature Range: RT to 50 °C, ±0.3°C in chamber feedback mode (sensor in heated glass lid and in chamber body)
•4.3'' touch screen interface
•Chambers available for any XY stage / piezo insert with Embedded temperature sensor in heated glass lid and in chamber body
•Compatible with: OBJ-COLLAR series, recommended for immersion objectives
•Pre-mixed gas flow rate can be regulated manually in the range 0.1-0.4 L/min or CO2/Air gas mixer
•Air Pump



1. Common microscope stages including Leica, Zeiss, Olympus, Marzhauser, ASI, Ludl & Prior:
1.1 H301-K-FRAME: heated chamber for inverted microscope stages with k-frame insert (160x110mm). Compatible with the following XY stages: Marzhauser (all models), Ludl (all models), ASI (all models, including z-piezo) Prior H107, Leica (k-frame), Zeiss (k-frame), Olympus (k-frame).
1.2. H301-MINI: heated chamber to fit in any microscope multi-well plate holder. Also compatible with stage Physik Instrumente PI P-737 and Ludl 99A600.
1.3. H301-UP: heated chamber for Upright Microscope
2. Leica Super-Z models:
2.1 H301-EC-LG-BL: heated chamber for Leica Super Z Galvo Stage. The chamber fits in the Super Z galvo stage (SP5/SP8), by replacing the Leica specimen holder. It can be screwed on the glavo Stage. Total chamber weight: 170 g.
2.2 Leica SP8 DLS Light Sheet models:
Electrically heated chamber for Leica DLS set-up includes objective collars for objective lenses HC APO L 10x/0.3 W and HC FLUOTAR L 25x/0.95 W uniformly heated by means of embedded electric resistances and is covered with a temperature controlled lid. Perfusion holes: #12 perfusion holes are available (2.0mm O.D.). The chamber fits in the Super Z galvo stage, by replacing the Leica specimen holder. It can be screwed on the galvo Stage. Magnetic specimen container holder for 35mm & 60mm petri dish.

2.3 H301-EC-LG-UP-BL: heated chamber for Upright Leica Super Z Galvo Stage. The chamber fits in the Super Z galvo stage (SP5/SP8), by replacing the Leica specimen holder. It can be screwed on the galvo Stage. Total chamber weight: 170 g.
3. Nikon microscope models
3.1 H301-NIKON-NZ100/200/500-N: heated chamber for Mad City Labs NZ100/200/500-N Z-piezo stages mounted on Nikon mechanical or motorized stages, TI-S-E and TI-S-ER.
3.2 H301-NIKON-TI-S-ER: heated chamber for Nikon Motorized XY Stage Ti-S-E or Ti-S-ER. Adapter is needed with Mad City Labs Piezo Stages NZ100-N/NZ200-N/NZ500-N mounted on Nikon motorized or mechanical stages.
3.3 H301-NIKON-TI-SR: heated chamber for Nikon T/Ti-SR rectangular stage with circular opening.
4. Olympus microscope models:
4.1 H301-OLYMPUS-IX-SUSP: heated chamber for Olympus rectangular XY stages with circular opening including IX-SUSP, GX-SVR, GX-SFR, BX3-SSU, IX2-SFR, IX2-SVL-2.
4.2 H301-OLYMPUS-IX3-SVR: heated chamber for Olympus IX3-SVR rectangular XY stage.
5. Prior microscope stage:
5.1 H301-PRIOR-H117: heated chamber for PRIOR H117 stage.
5.2 H301-PRIOR-NZ100-H107: heated chamber for Prior Z-stage NZ100/200 on Prior H107.
5.3 H301-PRIOR-NZ100-H117: heated chamber for Prior Z-stage NZ100/200 on Prior H117.
5.4 H301-PRIOR-NZ500: heated chamber for Prior NZ250/400/500 piezo Z-stages.
6.Physik Instrumente piezo stages
6.1 H301-PI-P-736-ZR1S/ZR2S:heated chamber for Physik Instrumente PI P-736-ZR1S/ZR2S.
6.2 H301-PI-736-160x110: heated chamber for Physik Instrumente PI P-736 with 160x110 aperture.
7. Ludl stage:
7.1 H301-LUDL-96A602: heated chamber for Ludl Piezo 96A602
8. Mad City Labs piezo stages
8.1 H301-MCL-Z100/500: heated chamber for Mad City Labs Z100/Z500 piezo Z-stages.
Holders for all types of specimen containers in different heating chamber:

Optional CO2 and O2 gas control:
3GF Mixer-Hypoxia 3 gas mixer (Carbon Dioxide, Nitrogen, Air) flowmeters control:
•CO2 range: 0-15%
•O2 range: 2-20%
•Set point resolution: 1%
•Filtering device: PTFE membrane with 0.2 mm pores
3GF Mixer-Hyperxia 3 gas mixer (Carbon Dioxide, Oxygen, Air) flowmeters control:
•CO2 range: 0-15%
•O2 range: 40-85%
•Set point resolution: 1%
•Filtering device: PTFE membrane with 0.2 mm pores

Optional Cooling and Heating Chamber & Control:
H101-CRYO-BL
•Temperature Range: from -5°C to 60°C
•Can operate in sample or chamber feedback mode
•Temperature Accuracy on sample: ±0.1°C in sample feedback mode; ±0.3°C in chamber feedback mode, if room T remains within 1°C
•Water Jacket Objective Heater